Dimension Standards

*

:::
Calibratd Device

(1) Ring Gauge
(2) Pin Gauge
(3) Plug Gauge 

Range (1) Ring Gauge: 1 mm~200 mm
(2) Pin Gauge: 1 mm~20 mm
(3) Plug Gauge: 20 mm~100 mm 
Uncertainty  (1) Ring Gauge: 2×[(0.132)2+(0.00136D)2]1/2 μm,D:Ring inner diameter in mm
(2) Pin Gauge: [(0.22)2+(0.013D)2]1/2μm,D:Pin gauge diameter size in mm
(3) Plug Gauge: 2×[(0.088)2+(0.00136D)2]1/2μm,D:Plug gauge diameter size in mm 
:::