【C07】Gas Measurement System

Calibrated Device  (1) Gas concentration detection tube, siren, leakage detector, gas concentration analyzer
(2) Gas Concentration Dilution Device (CO in N2、CO2 in N2、CH4 in Air、NO in N2、SO2 in N2)
Range (1)
CO:(0.0 to 0.1) mol/mol
CO2:(0 to 1) mol/mol
CH4:(0.00 to 0.05) mol/mol((0 to 100) %LEL)
C3H8:(0.00 to 0.02) mol/mol((0 to 100) %LEL)
(2)
CO in N2: (0 to 100) % @ (1 to 100) μmol/mol
CO2 in N2: (0 to 100) % @ (50 to 5000) μmol/mol
CH4 in Air: 0.10 to 100) % @ (1 to 20) mmol/mol
NO in N2: (0 to 100) % @ (1 to 2000) μmol/mol
SO2 in N2: (0 to 100) % @ (1 to 2000) μmol/mol
Uncertainty (1)
CO:2 μmol/mol
CO2:6 μmol/mol
CH4:59 μmol/mol(0.1 %LEL)
C3H8:26 μmol/mol(0.1 %LEL)
(2)
CO in N2: 0.02 %
CO2 in N2: 0.03 %
CH4 in Air: 0.15 %
NO in N2: 0.11 % @ (1 to 50) μmol/mol, 0.26 % @ (50 to 2000) μmol/mol
SO2 in N2: 0.048 % @ (1 to 50) μmol/mol, 0.13 % @ (50 to 2000) μmol/mol