Skip to main content

【U02】Microwave S-parameter Measurement System

Calibrated Device   Air Line, Open Circuit, Short Circuit, Sliding Short Circuit, Load, Sliding Load, Mismatch, Coaxial Line, Waveguide, Attenuator, Network Analyzer (Scattering Parameter Measurement)
Range 
Reflection coefficient 0 to 1 / (0 to 180°) (phase) /
Transmission coefficient (-60 to 0) dB / (0 to 180°) (phase)
  1. Frequency:10 MHz to 18 GHz(@Type N)
  2. Frequency:10 MHz to 26.5 GHz(@3.5 mm)
  3. Frequency:45 MHz to 40 GHz(@2.92 mm)
Uncertainty 
(1) Reflection (0.0054 to 0.0058) (linear) / 1.1° (phase)
Transmission (0.025 to 0.059) dB / 0.4° to 1.6° (phase)
(2) Reflection (0.0048 to 0.0084) (linear) / 2.3° to 2.4° (phase)
Transmission (0.068 to 0.28) dB / 0.5° to 1.8° (phase)
(3) Reflection (0.009 to 0.018) (linear) / 1.6° to 2.9° (phase)
Transmission (0.047 to 0.13) dB / 0.8° to 5.2° (phase)
Tags:
  • Last Updated:2020/02/26
Articles View Hits
9,833,752

A ccessibility