【U02】Microwave S-parameter Measurement System

Calibrated Device  

Air Line, Open Circuit, Short Circuit, Sliding Short Circuit, Load, Sliding Load, Mismatch, Coaxial Line, Attenuator

Range 
Reflection coefficient 0 to 1 / (0 to 180°) (phase) /
Transmission coefficient (-60 to 0) dB / (0 to 180°) (phase)
(1) Frequency:10 MHz to 18 GHz(@Type N)
(2) Frequency:10 MHz to 26.5 GHz(@3.5 mm)
(3) Frequency:45 MHz to 40 GHz(@2.92 mm)
Uncertainty 

(1) Reflection (0.0054 to 0.0058) (linear) / 1.1° (phase)               
     Transmission (0.025 to 0.059) dB / 0.4° to 1.6° (phase)
(2) Reflection (0.0048 to 0.0084) (linear) / 2.3° to 2.4° (phase)
     Transmission (0.068 to 0.28) dB / 0.5° to 1.8° (phase)
(3) Reflection (0.009 to 0.018) (linear) / 1.6° to 2.9° (phase)
     Transmission (0.047 to 0.13) dB / 0.8° to 5.2° (phase)