Skip to main content

【U02】Microwave S-parameter Measurement System (Service for High Frequency Dielectric Material will be resumed on 2021/01/04)

Calibrated Device  

(1) Air Line, Open Circuit, Short Circuit, Sliding Short Circuit, Load,
      Sliding Load, Mismatch, Coaxial Line, Attenuator

(2) High Frequency Dielectric Material

Range 
(1) Reflection coefficient 0 to 1 / (0 to 180°) (phase) /
      Transmission coefficient (-60 to 0) dB / (0 to 180°) (phase)
    (1.1) Frequency:10 MHz to 18 GHz(@Type N)
    (1.2) Frequency:10 MHz to 26.5 GHz(@3.5 mm)
    (1.3) Frequency:45 MHz to 40 GHz(@2.92 mm)
 
(2) Frequency:100 MHz ~ 26.5 GHz, Permittivity εr:1 ~ 50,
      Dielectric loss tan(δ) :0.0001 ~ 0.01
Uncertainty 

(1)
(1.1) Reflection (0.0054 to 0.0058) (linear) / 1.1° (phase)               
         Transmission (0.025 to 0.059) dB / 0.4° to 1.6° (phase)
(1.2) Reflection (0.0048 to 0.0084) (linear) / 2.3° to 2.4° (phase)
         Transmission (0.068 to 0.28) dB / 0.5° to 1.8° (phase)
(1.3) Reflection (0.009 to 0.018) (linear) / 1.6° to 2.9° (phase)
         Transmission (0.047 to 0.13) dB / 0.8° to 5.2° (phase)

(2) 0.15 % (relatively)