【O06】Absolute Radiometer System

Calibrated Device
(1) Luminous Intensity Standard Lamp
(2) IIIuminance Meter
(3) Chroma Meter
(4) Optical Detector
(5) Laser Light Source
Range
(1) 70 cd to 10000 cd(Candela Standard); 25 cd to 90000 cd
(2) 25 lx to 1500 lx
(3) IIIuminance: 25 lx to 1500 lx, Chromaticity coordinates (x , y): (0, 0) to (0.9, 0.9), (u , v): (0, 0) to (0.9, 0.9), Correlated Color Temperature: 2500 K to 3200 K
(4) Radiant Power: 6 μW to 100 mW, Wavelength: 300 nm to 9000 nm, IIIuminance Absolute Responsivity: 70 lx to 10000 lx
(5) Radiant Power: 50 μW to 150 mW, Irradiance: 50 μW/cm2 to 150 mW/cm2, Wavelength: 250 nm to 3000 nm
Uncertainty 
(1) 1.8 % (relative)(Candela Standard); 1.1 % to 1.5 % (relative), depending on luminous intensity
(2) 1.1 % ~ 1.5 % (relative), depending on IIIuminance range
(3) IIIuminance: 1.1 % ~ 1.5 % (relative), depending on IIIuminance range;  Chromaticity coordinates (x , y): (0.0012, 0.0008), (u , v): (0.0008, 0.0003); Correlated Color Temperature: 29 K
(4) Radiant Power Responsivity: 0.32 % to 0.54 %(relative), depending on wavelength range; Radiant Power: 0.28 % to 0.52 %(relative), depending on wavelength range; IIIuminance Absolute Responsivity: 1.8 %(relative)
(5) Radiant Power: 4.1 %(relative); Irradiance: 3.0 % to 6.2 %(relative) , depending on wavelength range