Calibrated Device
|
(1) Standard Color Plate, Filter
(2) Reflectance Standard
|
---|---|
Range
|
(1) Color Plate (0º:de/0º:di) Spectral Reflectance Factor:0.01 to 1, Wavelength:380 nm to 780 nm, Luminance factor:1 to 100, Chromaticity coordinates (x, y):(0, 0) to (1, 1), CIELAB L*:1 to 100, CIELAB (a*, b*):(-500, -200) to (500, 200) (de:8º/di:8º) Spectral Reflectance Factor:0.01 to 1, Wavelength:400 nm to 750 nm, Luminance factor:1 to 100, Chromaticity coordinates (x, y):(0, 0) to (1, 1), CIELAB L*:1 to 100, CIELAB (a*, b*):(-500, -200) to (500, 200) (0º:45ºa) Spectral Radiance Factor:> 0.01,Wavelength:380 nm to 780 nm, Luminance factor:> 1, Chromaticity coordinates (x, y):(0, 0) to (1, 1), CIELAB L*:> 1, CIELAB (a*, b*):(-500, -200) to (500, 200) Filter Spectral Transmittance: (1 to 100) %, Wavelength:380 nm to 800 nm, Luminous Transmittance: (1 to 100) % (2) Reflectance: (1 to 100) %, Wavelength: 250 nm to 2500 nm |
Uncertainty
|
(1) White Standard Spectral Radiance Factor: 0.013; Spectral Reflectance Factor: 0.0034 to 0.0042, depending on geometry; Luminance factor: 0.13 to 0.26, depending on geometry; Chromaticity coordinates (x, y): (0.0004, 0.0004) to (0.004, 0.004) , depending on geometry; CIELAB L* : 0.12 to 0.16, depending on geometry; CIELAB (a*, b*): (0.08, 0.06) to ( 0.30, 0.26), depending on geometry. Color Plate Luminance factor:0.17 to 0.29, depending on color; Chromaticity coordinates (x, y):(0.004, 0.004), depending on geometry and color; CIELAB L*:0.17 to 0.36,0.27 to 0.57, depending on geometry and color; CIELAB (a* , b*):(0.40 to 1.0, 0.32 to 1.2),(0.79 to 1.6, 0.64 to 1.2) , depending on geometry and color Filter Spectral Transmittance: 0.06 % to 0.21 %, depending on range; Luminous Transmittance: 0.04 % to 0.12 %, depending on range (2) Reflectance: 0.14 % |